The phantom for high resolution NM and PET systems performance evaluation (collimator, artifacts, calibration, reconstruction parameters). Among other things, it can be used to evaluate: center-of-rotation error, non-uniformity artifacts, changes of radius-of-rotation on spatial resolution, reconstruction filters on spatial resolution, attenuation and scatter compensation.
Technical data (can be modified to customer specifications):
Main cylinder
- inside cylinder diameter: 206 mm
- inside cylinder height: 186 mm
- cylinder wall thickness: 7 mm
Cold rods insert
- rod diameters: 3.2, 4.8, 6.4, 7.9, 9.5 and 11.1 mm
- height of rods: 88 mm
Cold spheres
- solid sphere diameters: 9.5, 12.7, 15.9, 19.1, 25.4 and 31.8 mm
- height of the center of the spheres from the base plate: 127 mm
Optional PET Lid with cylindrical samples
- refillable thin-walled cylinders, diameters: 8, 12, 16 and 25 mm
- water filled cylinder diameter: 25 mm
- air filled cylinder diameter: 25 mm
- PTFE solid cylinder diameter: 25 mm
- cylinder height: 38 mm
Accessoriess
- optional heavy duty carrying case
Product features:
- complies with:
- NEMA Standards Publication (NU 1-2001) Performance Measurements of Scintillation Cameras
- AAPM Report No. 9 - Computer Aided Scintillation Camera Acceptance Testing
- AAPM Report No. 22 - Rotating Scintillation Camera SPECT Acceptance Testing and Quality Control
- ACR–SNM (Res. 5 – 2011) technical standard for diagnostic procedures using radiopharmaceuticals - CE certified
- the manual provides detailed guidelines for carrying out each test, results assessment and registration