Pro-MAM Accreditation

03-501

The phantom can test the performance of a mammographic system and its ability to image small structures similar to those found clinically: calcifications, fibrous calcifications in ducts and tumor masses.

Technical data (can be modified to customer specifications):

  • dimensions: 101.5 x 108.0 x 44.0 mm
  • simulates 42 mm compressed breast of average glandular/adipose composition
  • total PMMA thickness: 36.75 mm
  • wax insert:
    – thickness: 7.25 mm
    – nylon fibrils diameters: 1.56, 1.12, 0.89, 0.75, 0.54 and 0.40 mm
    – microcalcifications: 0.54, 0.40, 0.32, 0.24 and 0.16 mm Al₂O₃ specks
    – tumor-like masses: 2.00, l.00, 0.75, 0.50 and 0.25 mm thick
  • 4 mm thick PMMA contrast disc
  • comfortable carrying case
Pro-MAM Accreditation pattern

Product features:

  • ACR Accredited
  • complies with:
    – IEC 61223-3-2
    – ACR Mammography Accreditation Program Requirements (May 4, 2012)
  • CE certified
  • the manual provides detailed guidelines for carrying out each test, results assessment and registration

Ask a Distributor for an offer

Locate a local distributor and access exclusive medical offers tailored to your needs.

Share this product
Close Cookie Preference Manager
Cookie Settings
By clicking “Accept All Cookies”, you agree to the storing of cookies on your device to enhance site navigation, analyze site usage and assist in our marketing efforts. More info
Strictly Necessary (Always Active)
Cookies required to enable basic website functionality.
Made by Flinch 77
Oops! Something went wrong while submitting the form.