Close iconSearch icon

Pro-RF AAPM 15

02-117

No items found.

This Digital Subtraction Angiography (DSA) Phantom follows the recommendation in Report No. 15 by the AAPM (American Associationof Physicists in Medicine) - Digital Radiology / Fluorography Task Group of the Diagnostic X-Ray Imaging Committee. Phantom is designed toevaluate digital functions of DSA systems and can be used to check: contrast range, resolution, linearity, uniformity, amplifier dynamic range,registration accuracy and subtraction effectiveness. Its modular construction allows setting up the desired test configuration in a very easy andaccurate manner.

Technical data (can be modified to customer specifications):

  • Section I and II
Pro-RF AAPM 15 Section 1 and 2

Step Wedge module, 6 step wedges (each 25.4 mm high) to test dynamic range. Can be folded into 203.2 x 203.2 x 76.5 mm solid block.

  • Section III
Pro-RF AAPM 15 Section 3

Slot block with a space to fit other test modules.
Outer size: 203.2 x 203.2 x 76.5 mm.
Slot size: 203.2 x 152.4 x 25.4 mm.

  • Registration plate
Pro-RF AAPM 15 Registration plate

Made of aluminium with an array of 3.18 mm holes.
Size: 203.2 x 203.2 x 1.5 mm

  • Bone section
Bone section

With three PTFE strips 25 mm in width and 5, 10 and 15 mm in thickness.
Size: 203.2 x 203.2 x 25.4 mm

  • High Contrast Resolution Section
Pro-RF AAPM 15 High Contrast Resolution Section

Containing pattern for line pair resolution evaluation (from 0.6 to 5.0 LP/mm).
Thanks to the dedicated recesses, test pattern can be placed in different orientations.
Size: 203.2 x 203.2 x 25.4 mm

  • Blank insert
Blank insert

Used with other modules forms a mask plate.
Size: 203.2 x 152.4 x 25.4 mm

  • Linearity Insert
Pro-RF AAPM 15 Linearity Insert

Contains 6 iodine areas (19 mm diameter) of different iodine thickness and concentration: 0.5, 1.0, 2.0, 4.0, 10.0 and 20.0 mg/cm².
Size: 203.2 x 152.4 x 25.4 mm

  • Low Contrast Artery Insert
Pro-RF AAPM 15 Low Contrast Artery Insert

Containing 3 sets of analogue blood vessel group (diameters 0.5, 1.0, 2,0 and 4.0 mm), each vessels group includes different iodine concentrations: 2.5, 5.0 and 10.0 mg/cc.
Size: 203.2 x 152.4 x 25.4 mm

  • 150 mg/ml Iodine Artery Insert
Pro-RF AAPM 15 150mg/ml Iodine Artery Insert

3 sets of analogue blood arteries (thickness 1.0, 2.0, 4.0 mm) each includes arteriarctial and arterial aneurysm whose sizes are 1/4, 1/2 and 3/4 of artery diameter.
Size: 203.2 x 152.4 x 25.4 mm

  • 300 mg/ml Iodine Artery Insert
Pro-RF AAPM 15 300mg/ml Iodine Artery Insert

3 sets of analogue blood arteries (thickness 1.0, 2.0, 4.0 mm) each includes arteriarctia and arterial aneurysm whose sizes are 1/4, 1/2 and 3/4 of artery diameter.
Size: 203.2 x 152.4 x 25.4 mm

  • Carrying Case

Product features:

  • complies with:
    – AAPM Report No. 15 - Performance Evaluation and Quality Assurance in Digital Subtraction Angiography - Diagnostic X-Ray Imaging Committee Digital Radiography / Fluorography Task Group - May 1985
    – IEC 61223-3-1
  • CE certified
  • the manual provides detailed guidelines for carrying out each test, results assessment and registration
Share
Close Cookie Preference Manager
Cookie Settings
By clicking “Accept All Cookies”, you agree to the storing of cookies on your device to enhance site navigation, analyze site usage and assist in our marketing efforts. More info
Strictly Necessary (Always Active)
Cookies required to enable basic website functionality.
Made by Flinch 77