The phantom for evaluation of imaging performance of radiographic systems. It allows wide-range low-contrast and high-contrastresolution evaluation and sensitometric measurements. Available in standard layout plus any custom configuration.

Technical data (can be modified to customer specifications):

  • diameter: 180 mm
  • thickness: 10 mm
  • 3 circular rims on the outside of the phantom (Ø: 150, 160 and 170 mm) for geometry and collimation check
  • 17 low-contrast objects, 11 mm in diameter, producing contrast values from 0.2 to 7.5%
  • 17 small high-contrast objects, 0.5 mm in diameter, producing contrast values from 3.9 to 95.4%
  • 10 sensitometric measurement point details, 5.6 in diameter, producing contrast values from 0.11 to 1.0
  • pattern for line pair resolution evaluation (from 0.5 to 14.3 LP/mm)
  • 1.0 and 0.5 mm high purity Cu attenuation filters (fits to the diaphragm housing)
  • compact carrying case

Product features:

  • complies with:
    – IEC 61223-3-1
    – IPEM Report 91 (2005): Recommended Standards for the Routine Performance Testing of Diagnostic X-Ray Imaging Systems published by the Institute of Physics and Engineering in Medicine
  • CE certified
  • the manual provides detailed guidelines for carrying out each test, results assessment and registration

Ask a Distributor for an offer

Locate a local distributor and access exclusive medical offers tailored to your needs.

Share this product
Close Cookie Preference Manager
Cookie Settings
By clicking “Accept All Cookies”, you agree to the storing of cookies on your device to enhance site navigation, analyze site usage and assist in our marketing efforts. More info
Strictly Necessary (Always Active)
Cookies required to enable basic website functionality.
Made by Flinch 77
Oops! Something went wrong while submitting the form.