The phantom for performance evaluation of small animal systems or ultra-high resolution ECT systems. It can be used to evaluate: center-of-rotation error, non-uniformity artifacts and spatial resolution. Available in two sizes for FOV at least 45 mm or 75 mm.
Technical data (can be modified to customer specifications):
At least 75 mm FOV version:
Main cylinder:
- outside cylinder diameter: 80 mm
- inside cylinder diameter: 74 mm
- inside cylinder height: 65 mm
Cold rods insert:
- rod diameters: 1.2, 1.6, 2.4, 3.2, 4.0 and 4.8 mm
- height of rods: 34 mm
At least 45 mm FOV version:
Main cylinder:
- outside cylinder diameter: 50 mm
- inside cylinder diameter: 45 mm
- inside cylinder height: 65 mm
Cold rods insert:
- rod diameters: 1.2, 1.6, 2.4, 3.2, 4.0 and 4.8 mm
- height of rods: 34 mm
Accessories
Product features:
- complies with:
- NEMA Standards Publication (NU 1-2001) Performance Measurements of Scintillation Cameras
- AAPM Report No. 9 - Computer Aided Scintillation Camera Acceptance Testing
- AAPM Report No. 22 - Rotating Scintillation Camera SPECT Acceptance Testing and Quality Control
- ACR–SNM (Res. 5 – 2011) technical standard for diagnostic procedures using radiopharmaceuticals - CE certified
- the manual provides detailed guidelines for carrying out each test, results assessment and registration